Overview
Fluke 1621 overview
- 3-pole Fall-of-Potential earth testing for basic measurements
- 2-pole resistance measurements for added versatility
- Easily capture values with single-button operation
- Ensure accurate measurements with automatic voltage detection
- Hazardous voltage warning offers increased user protection
- Clearly read and record data with a large, backlit display
- Rugged holster and design for tough work environments
- Portable size allows for easy transportation
- Instantly be alerted to measurements outside of your set limit, when you use the adjustable limit setting
- 600 V Cat II
|
|
Description
Fluke 1621 data sheet Click to download Fluke 1621 datasheet.
The Fluke 1621 is an easy-to-use earth ground tester. For ground resistance testing, the 1621 is the first line of defense in detecting reliable ground connections. The unit features basic ground testing methods including 3-pole Fall-of-Potential as well as 2-pole ground resistance tests. Its convenient size, rugged holster, and large, clear LCD display make it an ideal field earth ground tester, for most electrical grounding work environments. With a simple user interface and intuitive functionality, the Fluke 1621 is a handy grounding tool for electrical contractors, utility test engineers, and earth ground specialists.
What's Included?
- Users manual
- Two measuring leads with alligator clips, 2 m (6 ft)
- One battery, 9 V alkaline (LR61)
- One protective holster, yellow
- One CD-ROM
NSN 6625015712654
Specs relating to Fluke 1621 Earth Ground Tester
| General specifications |
| Measuring functions |
|
3-pole earth ground resistance, 2-pole ac resistance of a conductor, Interference voltage |
|
| Intrinsic error |
|
Refers to the reference temperature range and is guaranteed for one year |
|
| Measuring rate |
|
| Battery |
|
| Battery condition |
|
LO-BAT is displayed if voltage drops below 6.5 V |
|
| Voltages |
| Between jacks H/C2 and E/C1: |
250 Veff maximum (effective voltage) |
| Between jacks S/P2 and E/C1: |
250 Veff maximum |
|
| Climatic class |
|
VDE/VDI 3540 RZ (conforming to KWG as per DIN 40040, 4/87) |
|
| Temperature performance[2] |
| Working: |
-10 °C to +50 °C |
| Operating: |
0 °C to +35 °C |
| Storage: |
-20 °C to +60 °C |
| Reference: |
+23 °C ± 2 °C |
|
| Temperature coefficient |
|
± 0.1 % of range per degree Kelvin |
|
| Safety |
|
IEC/EN 61010-1, 600 V CAT II |
|
| Dimensions |
|
113 mm x 54 mm x 216 mm including holster |
|
| Weight |
|
850 g, including standard accessories, volume approximately 600 cm3 |
|
| Note: |
|
[1]If the tester is not going to be used, or is being stored for a long period, remove the battery and store separately from the tester to avoid damage from battery leakage. |
|
| Note: |
|
[2]The four temperature ranges for the tester exists to satisfy European Standards requirements; the instrument can be used over the full working temperature range by using the temperature coefficient to calculate accuracy at the ambient temperature of use. |
|
| Electrical specifications |
| Maximum deviations: |
| E1 Influence factor |
Position |
| E1 Deviation influence |
0 % |
| E2 Influence factor |
Supply voltage |
| E2 Deviation influence |
0 % |
| E3 Influence factor |
Temperature E3 |
| E3 Deviation influence |
2.3 % |
| E4 Influence factor |
Serial interference voltage (20 V) |
| E4 Deviation influence |
0.6 % |
| E5 Influence factor |
Probe- and auxiliary probe resistance |
| E5 Deviation influence |
10 % |
|
| Test voltage |
|
| Protection type |
|
| Electromagnetic compatibility |
|
Emission: IEC/EN 61326 Class B |
|
Immunity: IEC/EN 61326 Annex C |
|
| RE resistance measurement |
| Measuring method |
Current-voltage measurement with improved cross-talk attenuation, no compensation of measuring lead resistance, with probe (3-pole) or without probe (2-pole), as per IEC/EN 61557-5 |
| Open circuit voltage |
23 to 24 V ac |
| Short circuit current |
> 50 mA ac |
| Measuring frequency |
128 Hz |
| Maximum permissible overload |
250 Veff |
|
| Measuring time |
|
8 seconds (average from when START is pressed) |
|
| Limit input |
|
Tester retains set value even if instrument is turned off (assuming battery power supply is sufficient) |
|
| Automatic changeover of resolution |
| RH |
< 7 kΩ |
| Resolution |
0.01 Ω |
| RH |
< 50 kΩ |
| Resolution |
0.1 Ω |
| RH |
> 50 kΩ |
| Resolution |
1 Ω |
|
| Interference voltage display dc + ac |
| Vmax |
30 Veff |
| Common mode rejection |
> 80 dB at 50 Hz and 60 Hz |
| Ri |
680 kΩ |
| Measuring uncertainty |
< 10 % for pure ac and dc signals |
|
| Measuring range |
| 0.15 Ω to 20 Ω |
| Resolution |
0.01 Ω |
| Display range |
0 to 19.99 Ω |
|
| 200 Ω |
| Resolution |
0.1 Ω |
| Display range |
20 to 199.9 Ω |
|
| 2 kΩ |
| Resolution |
1 Ω |
| Display range |
200 to 1999 Ω |
|
| Intrinsic uncertainty |
| |
± (6 % of measured value + 5D) |
|
| Operating uncertainty IEC 61557[1] |
| |
± (18 % of measured value + 5D) |
|
| Notes: |
| |
[1] Covers all deviations caused by influence quantities E1-E5. If the deviation E4 caused by high probe or auxiliary probe resistance is higher than specified flashes. Measured values are outside of the specified operating uncertainty. |
|
What's Included?
- Users manual
- Two measuring leads with alligator clips, 2 m
- One battery, 9 V alkaline (LR61)
- One protective holster, yellow
- One CD-ROM
|