- QWERTY keyboard for rapid data entry
- Compact Flash memory card capability for back-up data storage + transfer to PC
- Large backlit graphical display
- Pre-set, auto-test sequences for user convenience
- Integral site, location and description codes for faster data processing
- Memory review facility for more on-site control
Description
Fluke 6500 PAT Tester
This PAT tester is perfect for testing all class 1 and class 2 appliances.
Applications: Testing appliances, and leads in accordance with the AS/NZS 3760:2003. Built in memory records Site, Location, and asset details, in addition to the results of the test. Mains operated, this unit is highly portable but eqaually suited to bench use boasting a full QWERTY keyboard. Help screens are detailed assisting the user in testing procedures.
Test Specifications
Power-on Test
Display range:
90 V to 264 V
Accuracy at 50 Hz:
± (2% + 3 counts)
Resolution:
0.1 V (1 V - model 6200)
Input impedance
> 1 MΩ // 2.2 nF
Maximum input mains voltage:
300 V
Earth Bond Test (Rpe)
Display range:
0 to 19.99 Ω
Accuracy (after Bond Test zeroing):
± (2.5% + 4 counts)
Resolution:
0.01 Ω
Test current:
200 mA AC -0% +40% into 1.99 Ω
25 A AC ± 20% into 25 mΩ at 230 V
Open circuit voltage:
> 4 V AC, < 24 V AC
Bond test zeroing:
can subtract up to 1.99 Ω
Insulation Test (Riso)
Display range:
0 to 299 MΩ
Accuracy:
± (5% + 2 counts) from 0.1 to 300 MΩ
Resolution:
0.01 MΩ (0 to 19.99 MΩ)
0.1 MΩ (20 to 199.9 MΩ)
1 MΩ (200 to 299 MΩ)
Test voltage:
500 V DC -0% + 10% at 500 kΩ load
Test current:
> 1 mA at 500 kΩ load, < 15 mA at 0 Ω
Auto discharge time:
< 0.5 s for 1 μF
Max. capacitive load:
operational up to 1 μF
Touch Current Test
Display range:
0 to 1.99 mA AC
Accuracy:
± (4% + 2 counts)
Resolution:
0.01 mA
Internal resistance (via probe):
2 kΩ
Measuring method:
probe
The appliance under test is energized at mains potential.
Substitute Leakage Current Test
Display range:
0 to 19.99 mA AC
Accuracy:
± (2.5% + 3 counts)
Resolution:
0.01 mA
Test voltage:
35 V AC ± 20%
Load/Leakage Test: Load Current
Display range:
0 to 13 A
Accuracy:
± (4% + 2 counts)
Resolution:
0.1 A
The appliance under test is energized at mains potential.
Load/Leakage Test: Load Power
Display range:
0 to 999 VA
1.0 kVA to 3.2 kVA
Accuracy:
± (5% + 3 counts)
Resolution:
1 VA (0 to 999 VA)
0.1 kVA (1.0 kVA to 3.2 kVA)
The appliance under test is energized at mains potential.
Load/Leakage Test: Leakage Current
Display range:
0 to 19.99 mA
Accuracy:
± (4% + 4 counts)
Resolution:
0.01 mA
The appliance under test is energized at mains potential.
PELV Test
Accuracy at 50 Hz:
± (2% + 3 counts)
Overload protection:
300 Vrms
Warning threshold:
25 Vrms
*The measurement range meets the service operating errors specified in EN61557-1: 1997, EN61557-2: 1997 , EN61557-4: 1997
*The accuracy specifications above for the display range is defined as ± (%reading + digit counts) at 23 °C ± 5 °C, ≤ 75 % RH. Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy specification) per °C.
Included accessories: Earth lead, mains lead, certificate of conformity, hard carry case.